Pharma Climatic Test Chambers, Series CP
For long-term stability test with highest testing precision.

Standard version of the product line "CP": |
 |
capacitive humidity measuring system |
 |
casing in stainless-steel |
 |
door lockable |
 |
entry port ø 50 mm |
 |
6 shelves |
 |
CTS-Terminal |
 |
low noise refrigerating unit |
 |
potential-free contact for malfunction signal |
 |
interface RS 232 |
Your benefits:
|
 |
use of environmentally friendly materials and refrigerants |
 |
own CTS-controller optimized to CTS-systems |
 |
simple user-friendly operation |
 |
internal and external material in stainless-steel for easy cleaning |
Options:
|
 |
door with window |
 |
validation of the CTS unit (IQ, OQ) |
 |
Software validation |
Test space capacity in liters:
|
 |
500 / 1100 |
Temperature range:
|
 |
+10 °C / +50 °C |
Humidity range:
|
 |
10 - 90 % RH |
Dew Point range I:
|
 |
+7 °C / +45 °C |
Dew Point range II:
|
 |
-5 °C / +7 °C |
Design:
|
| Control: |
microprocessor control and monitoring system |
| Test space: |
stainless steel grade 1.4301 |
| Casing: |
stainless steel grade 1.4301 |
| Refrigerants: |
chloride free, hermetically sealed |
| Humidity system: |
capacitive measuring |
Options:
|
 |
add. entry ports ø 50 mm |
 |
add. shelves |
 |
add. Pt 100 for temperature measuring on test specimen alternative reversible to controller function |
 |
door with full-size window (reduced dew-point-range < +30 °C) |
 |
extended temperature range down to ± 0 °C |
 |
reservoir for humidification water |
 |
temperature protection for test-specimen |
 |
Touch Panel, colour display LCD |
 |
CID software for programming and documentation |
 |
RS 232 interface changeable to USB |
 |
RS 485 |
| Type |
CP 10/500 |
CP 10/1100 |
| Test Space |
liters |
500 |
1100 |
| Test Space Dimensions |
height |
mm, appr. |
1225 |
1225 |
| width |
mm, appr. |
595 |
2 x 595 = 1190 |
| depth |
mm, appr. |
620 |
620 |
| Overall Dimensions |
height |
mm, appr. |
2080 |
2080 |
| width |
mm, appr. |
720 |
2 x 720 = 1440 |
| depth |
mm, appr. |
850 |
850 |
| |
| Temperature Tests |
| Temperature Range |
°C |
+10 / +50 |
| Temperature Fluctuation |
K |
± 0,5 temporally |
| |
| Climatic Tests |
| Temperature Range |
°C |
+10 / +50 |
| Humidity Range |
% RH |
10 / 90 |
| Dew Point Range I |
°C |
+7 / +45 |
| Dew Point Range II |
°C |
-5 / +7 |
| Humidity Fluctuation |
% RH |
± 1 / 3 |
| Temperature Fluctuation |
K |
± 0,5 temporally |
| |
| Nominal Voltage |
230 V +6% -10%, 1/N, 50 Hz |
| Nominal Power |
kW, appr. |
1,6 |
2,0 |
| Condenser |
|
air-cooled |
| Weight |
kg, appr. |
150 |
200 |
All figures are average values which have been obtained at a temperature of +25° C, without test specimens, without thermal load and without options.
We reserve the right to make alterations due to technical development.